The Use of Two DPM Modes Equipped with a Thermohygrometer and Pressure Selection for the Implementation of Community Service

Authors

  • Syaifudin Syaifudin Department of Electromedical Engineering, Poltekkes Kemenkes Surabaya
  • Ketut Dyah Kusumawati Department of Electromedical Engineering, Poltekkes Kemenkes Surabaya
  • Tribowo Indrato Department of Electromedical Engineering, Poltekkes Kemenkes Surabaya

DOI:

https://doi.org/10.35882/ficse.v1i1.5

Keywords:

Calibration, tensimeter, humidity, community health center, community service.

Abstract

Calibration is a technical activity consisting of determination, determination of one or more properties and characteristics of a product, process,s or service in accordance with a specific procedure that has been determined. The purpose of calibration is to ensure that the measurement results comply with national and international standards. Digital Pressure Meter (DPM) Two Modes equipped with Thermohygrometer and Pressure Selection is a device used to determine positive pressure and negative pressure (vacuum). Using the MPX 5050GP sensor as a positive pressure sensor. maximum pressure requirement is 300 mmHg. Using a DHT 22 sensor for temperature humidity. This device is also equipped with a leak test timer. The display used in this module is a 2.4-inch TFT Nextion LCD. After carrying out the measurement process to the mercury sphygmomanometer 6 times, the smallest result was 0 mmHg at 0 mmHg pressure and the largest was 298.0 mmHg at 300mmHg pressure. Based on the temperature humidity data collection carried out in the room obtained 31̊C at temperature and 87% at humidity.

References

Prasetyo Wicaksono. (2015). Vacum, T., & Positif. surabaya: POLTEKKES SURABAYA.

Rosyi Dwi Putranti. (2016). Analisis Perbandingan Tensimeter. surabaya: POLTEKKES SURABAYA.

M. Junia Dyah Permata Wibisono, Priyambada Cahya Nugraha, MT, Hj. Andjar Pudji, ST and ABSTRAK, “‘ Digital Pressure Meter ( DPM ) Va cum Pressure ,’” Jur. Tek. Elektromedik Politek. Kesehat. KEMENTRIAN Kesehat. SURABAYA, 2017.

S. T. Yosep KurAkhir, J. Teknik, E. Politeknik, and K. Surabaya, “Dpm dua mode,” 2018.

M. Ryan, N. Rokhman, B. G. Irianto, and H. G. Ariswati, “DIGITAL PRESSURE METER,” vol. 1, no. 1, pp. 1–4, 2019.

Turner, M. J., Speechly, C., & Bignell, N. (2007). Sphygmomanometer calibration Why, how and how often? Australian Family Physician, 36(10), 834–837.

You, A., Be, M. A. Y., & In, I. (2019). A simple calibration methods of relative humidity sensor DHT22 for tropical climates based on Arduino data acquisition system. 020009(January).

Vatsal, S., & Bhavin, M. (2017). Using Raspberry Pi To Sense Temprature and Relative Humidity. International Research Journal of Engineering and Technology(IRJET), 4(2), 380–385. https://irjet.net/archives/V4/i2/IRJET-V4I276.pdf

de Greeff, A., Lorde, I., Wilton, A., Seed, P., Coleman, A. J., & Shennan, A. H. (2010). Calibration accuracy of hospital-based non-invasive blood pressure measuring devices. Journal of Human Hypertension, 24(1), 58–63. https://doi.org/10.1038/jhh.2009.29

O’Brien, E., Waeber, B., Parati, G., Staessen, J., & Myers, M. G. (2001). Blood pressure measuring devices: Recommendations of the European Society of Hypertension. British Medical Journal, 322(7285), 531–536. https://doi.org/10.1136/bmj.322.7285.531

Avendaño, G., Fuentes, P., Castillo, V., Garcia, C., & Dominguez, N. (2010). Reliability and safety of medical equipment by use of calibration and certification instruments. LATW2010 - 11th Latin-American Test Workshop, 4–7. https://doi.org/10.1109/LATW.2010.5550349

Sihombing, Y. A., & Listiari, S. (2020). Detection of air temperature, humidity and soil pH by using DHT22 and pH sensor based Arduino nano microcontroller. AIP Conference Proceedings, 2221(March). https://doi.org/10.1063/5.0003115

Oo, A. Z. (2018). SERVER BASED REAL-TIME ENVIRONMENTAL. XVI(2).

NLA L, "Calibrator Tensimeter Equipped With PC-based Thermohygrometer," Tensim calibrator. Equipped with Thermohygrom. Berbas. PC Nov., P. 2, 2017.

Abdul Cholid Ridwan. (2019). DPM Two Modes Equipped with Temperature and Humidity. surabaya: POLTEKKES SURABAYA.

DW Wulandari, E. Swistoro, and C. Connie, "Effectiveness of aneroid sphygmomanometer modification as a measurement of hydrostatic pressure and its implementation as props," PENDIPA J. Sci. Educ., Vol. 2, no. 1, pp. 82-87, 2018.

A. Türk and A. Hamarat, "Automated Pressure Calibration of Blood Pressure Measuring Device Calibrator to Realize Its Traceability" Med. Meas. Appl. MeMeA 2019 - Symp. Proc., Pp. 1-5, 2019.

RF Muldiani and K. Hadiningrum, "Optimization Tool Practical Law of Thermodynamics," Pros. SNFA (Seminar Nas. Fis. And App., Vol. 3, pp. 237-245, 2018.

AH Noviyanto, "Pressure Sensor Applications Mpxm2053Gs In sphygmomanometer Pressure Test Tool Based Microcontroller ATmega328," vol. 21, no. 1, pp. 87-94, 2017.

S. Shin, D. Seo, M. Hwang, and S. Song, "A Study on the Verification of Terminology Classification," vol. 29, no. 13, pp. 2797-2799, 2011

Downloads

Published

2022-01-26

Issue

Section